Efficient determination of average valence of manganese in manganese oxides by reaction headspace gas chromatography.

Authors :
Gong, Yi-Xian
Xie, Wei-Qi
Yu, Kong-Xian
Source :
Journal of Chromatography A. Aug2017, Vol. 1511, p138-142. 5p.
Publication Year :
2017

Abstract

This work investigates a new reaction headspace gas chromatographic (HS-GC) technique for efficient quantifying average valence of manganese (Mn) in manganese oxides. This method is on the basis of the oxidation reaction between manganese oxides and sodium oxalate under the acidic condition. The carbon dioxide (CO 2 ) formed from the oxidation reaction can be quantitatively analyzed by headspace gas chromatography. The data showed that the reaction in the closed headspace vial can be completed in 20 min at 80 °C. The relative standard deviation of this reaction HS-GC method in the precision testing was within 1.08%, the relative differences between the new method and the reference method (titration method) were no more than 5.71%. The new HS-GC method is automated, efficient, and can be a reliable tool for the quantitative analysis of average valence of manganese in the manganese oxide related research and applications. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00219673
Volume :
1511
Database :
Academic Search Index
Journal :
Journal of Chromatography A
Publication Type :
Academic Journal
Accession number :
124143630
Full Text :
https://doi.org/10.1016/j.chroma.2017.06.061