Preanalytic process linked to spuriously elevated HIV viral loads: improvement on an FDA-approved process.

Authors :
Procop, Gary W.
Taege, Alan J.
Starkey, Colleen
Tungsiripat, Marisa
Warner, Diane
Schold, Jesse D.
Yen-Lieberman, Belinda
Source :
Diagnostic Microbiology & Infectious Disease. Sep2017, Vol. 89 Issue 1, p44-46. 3p.
Publication Year :
2017

Abstract

The processing of specimens often occurs in a central processing area within laboratories. We demonstrated that plasma centrifuged in the central laboratory but allowed to remain within the primary tube following centrifugation was associated with spuriously elevated HIV viral loads compared with recentrifugation of the plasma just prior to testing. [ABSTRACT FROM AUTHOR]