Back to Search Start Over

Variance components associated with long-echo-time MR spectroscopic imaging in human brain at 1.5T and 3T.

Authors :
Thrippleton, Michael J.
Parikh, Jehill P.
Semple, Scott I. K.
Harris, Bridget A.
Andrews, Peter J. D.
Wardlaw, Joanna M.
Marshall, Ian
Source :
PLoS ONE. 12/29/2017, Vol. 12 Issue 12, p1-11. 11p.
Publication Year :
2017

Abstract

Object: Magnetic resonance spectroscopic imaging (MRSI) is increasingly used in medicine and clinical research. Previous reliability studies have used small samples and focussed on limited aspects of variability; information regarding 1.5T versus 3T performance is lacking. The aim of the present work was to measure the inter-session, intra-session, inter-subject, within-brain and residual variance components using both 1.5T and 3T MR scanners. Materials and methods: Eleven healthy volunteers were invited for MRSI scanning on three occasions at both 1.5T and 3T, with four scans acquired at each visit. We measured variance components, correcting for grey matter and white matter content of voxels, of metabolite peak areas and peak area ratios. Results: Residual variance was in general the largest component at 1.5T (8.6–24.6%), while within-brain variation was the largest component at 3T (12.0–24.7%). Inter-subject variation was around 5%, while inter- and intra-session variance were both generally small. Conclusion: Multiple variance contributions associated with MRSI measurements were quantified and the performance of 1.5T and 3T MRI scanners compared using data from the same group of subjects. Residual error is much lower at 3T, but other variance components remain important. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
19326203
Volume :
12
Issue :
12
Database :
Academic Search Index
Journal :
PLoS ONE
Publication Type :
Academic Journal
Accession number :
127052406
Full Text :
https://doi.org/10.1371/journal.pone.0189872