Back to Search Start Over

A quantitative study on magnesium alloy stent biodegradation.

Authors :
Gao, Yuanming
Wang, Lizhen
Gu, Xuenan
Chu, Zhaowei
Guo, Meng
Fan, Yubo
Source :
Journal of Biomechanics. Jun2018, Vol. 74, p98-105. 8p.
Publication Year :
2018

Abstract

Insufficient scaffolding time in the process of rapid corrosion is the main problem of magnesium alloy stent (MAS). Finite element method had been used to investigate corrosion of MAS. However, related researches mostly described all elements suffered corrosion in view of one-dimensional corrosion. Multi-dimensional corrosions significantly influence mechanical integrity of MAS structures such as edges and corners. In this study, the effects of multi-dimensional corrosion were studied using experiment quantitatively, then a phenomenological corrosion model was developed to consider these effects. We implemented immersion test with magnesium alloy (AZ31B) cubes, which had different numbers of exposed surfaces to analyze differences of dimension. It was indicated that corrosion rates of cubes are almost proportional to their exposed-surface numbers, especially when pitting corrosions are not marked. The cubes also represented the hexahedron elements in simulation. In conclusion, corrosion rate of every element accelerates by increasing corrosion-surface numbers in multi-dimensional corrosion. The damage ratios among elements with the same size are proportional to the ratios of corrosion-surface numbers under uniform corrosion. The finite element simulation using proposed model provided more details of changes of morphology and mechanics in scaffolding time by removing 25.7% of elements of MAS. The proposed corrosion model reflected the effects of multi-dimension on corrosions. It would be used to predict degradation process of MAS quantitatively. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00219290
Volume :
74
Database :
Academic Search Index
Journal :
Journal of Biomechanics
Publication Type :
Academic Journal
Accession number :
129683026
Full Text :
https://doi.org/10.1016/j.jbiomech.2018.04.027