Back to Search Start Over

Circuitry Design and Magnetic Susceptibility Evaluation of 7T fMRI Implantable RF Coil.

Authors :
Rong Wang
Dong, Celia M.
Wu, Ed X.
Roberts, Robert C.
Li Jun Jiang
Source :
Applied Computational Electromagnetics Society Journal. Oct2018, Vol. 33 Issue 10, p1109-1111. 3p.
Publication Year :
2018

Abstract

Implantable coils have been widely utilized in functional magnetic resonance imaging (fMRI) owing to their superior signal to noise ratio. To effectively minimize the magnetic field distortion and image artifacts, the magnetic susceptibility of the implantable coil needs to be evaluated before practical use. In this work, we experimentally identify the magnetic susceptibility of each component of the implantable coil with the 7T Bruker NMR imaging scanner and provide useful guidelines for the following manufacturing. A 5 x 5 mm² implantable surface coil with a tunable frequency range is subsequently introduced for the 7T fMRI of the rat primary somatosensory cortex (S1FL). A detachable external tuning circuit for the implantable coil is employed to facilitate in-vivo measurements in the rat model. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
10544887
Volume :
33
Issue :
10
Database :
Academic Search Index
Journal :
Applied Computational Electromagnetics Society Journal
Publication Type :
Academic Journal
Accession number :
133741571