Back to Search Start Over

Design and preliminary test of a multi-element tissue-equivalent proportional counter based on the gas electron multiplier.

Authors :
Zhang, Weihua
Li, Chunjuan
Wang, Zhiqiang
Qi, Huirong
Zou, Yisheng
Liu, Yina
Luo, Hailong
Source :
Nuclear Instruments & Methods in Physics Research Section A. Feb2019, Vol. 916, p37-41. 5p.
Publication Year :
2019

Abstract

Abstract In order to monitor the intense neutron and gamma radiation, multi-element tissue equivalent proportional counters (TEPC) based on the standard gas electron multiplier (GEM) and ceramic thick GEM (THGEM) foils were designed and respectively tested by implanted 55Fe and 241Am sources which would also be used as calibration sources. The performance testing of GEM-TEPC/THGEM-TEPC detectors at various parameters showed that the detectors could work well. The THGEM-TEPC was also used to measure the linear energy spectra in 241Am–Be and intense (108 ncm−2s−1) neutron radiation fields. The results displayed that THGEM-TEPC was available to do microdosimetric research or be a dosimeter in intense neutron/gamma field. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
01689002
Volume :
916
Database :
Academic Search Index
Journal :
Nuclear Instruments & Methods in Physics Research Section A
Publication Type :
Academic Journal
Accession number :
133780865
Full Text :
https://doi.org/10.1016/j.nima.2018.11.006