Abstract

To be clinically viable, a brain-machine interface (BMI) requires transcutaneous telemetry. Spike-based compression algorithms can be used to reduce the amount of telemetered data, but this type of system is subject to queuing-based transmission delays. This paper examines the relationships between the ratio of output to average input bandwidth of an implanted device and transmission latency and required queue depth. The examination was performed with a computer model designed to simulate the telemetry link. The input to the model was presorted spike data taken from a macaque monkey performing a motor task. The model shows that when the output bandwidth/average input bandwidth is in unity, significant transmission latencies occur. For a 32-neuron system, transmitting 50 bytes of data per spike and with an average neuron firing rate of 8.93 spikes/s, the average maximum delay was approximately 3.2 s. It sis not until the output bandwidth is four times the average input bandwidth that average maximum delays are reduced to less than 10 ms. A comparison of neuron firing rate and resulting latencies shows that high latencies result from neuron bursting. These results will impact the design of transcutaneous telemetry in a BMI. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00189294
Volume :
51
Issue :
6
Database :
Academic Search Index
Journal :
IEEE Transactions on Biomedical Engineering
Publication Type :
Academic Journal
Accession number :
13400293
Full Text :
https://doi.org/10.1109/TBME.2004.827090