Calibration of a [formula omitted]Li diamond-sandwich spectrometer with quasi-monoenergetic neutrons.

Authors :
Osipenko, M.
Ceriale, V.
Gariano, G.
Girolami, M.
Cerchi, S.
Nolte, R.
Pirovano, E.
Ripani, M.
Trucchi, D.M.
Source :
Nuclear Instruments & Methods in Physics Research Section A. Jul2019, Vol. 931, p135-141. 7p.
Publication Year :
2019

Abstract

In the present article we describe calibration measurements of the compact neutron spectrometer at the PTB quasi-monoenergetic neutron reference fields. The spectrometer is based on two diamond detectors enclosing a 6 Li neutron converter. The conversion occurs through the 6 Li(n,t) 4 He reaction, featuring the Q -value of 4.8 MeV. The incident neutron energy is measured through the sum of energies of the two conversion reaction products: t and α. The prototype used in this work was based on two 4 × 4 mm2 100 μ m thick single crystal CVD commercial diamond plates. The measurements were performed in the fission spectrum domain at five different neutron energies varying from 0.3 up to 2 MeV. The obtained data showed a very good neutron energy reconstruction, i.e. in agreement with its reference values within statistical uncertainties or with absolute deviations below 11 keV. The energy resolution of the spectrometer was found to be about 100 keV (RMS) with a small (about 9 keV/MeV) rising trend towards higher energies. The measured detection efficiency was found to be compatible with predictions based on analytical calculations and on Geant4 simulations within 3% statistical and 4% systematic uncertainty. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
01689002
Volume :
931
Database :
Academic Search Index
Journal :
Nuclear Instruments & Methods in Physics Research Section A
Publication Type :
Academic Journal
Accession number :
136134808
Full Text :
https://doi.org/10.1016/j.nima.2019.04.015