Back to Search Start Over

Data Acquisition System for Thomson Scattering Diagnostics on GDT.

Authors :
Puryga, E. A.
Lizunov, A. A.
Ivanenko, S. V.
Khilchenko, A. D.
Kvashnin, A. N.
Zubarev, P. V.
Moiseev, D. V.
Source :
IEEE Transactions on Plasma Science. Jun2019, Vol. 47 Issue 6, p2883-2889. 7p.
Publication Year :
2019

Abstract

This paper describes a data acquisition system for Thomson scattering diagnostics on a gas-dynamic trap (GDT) (the Budker Institute of Nuclear Physics, Novosibirsk, Russia). It allows recording scattered laser radiation pulses of ~10-ns duration with 5-GSPS sampling rate and 14-bit amplitude resolution. The data acquisition system includes eight-channel high-speed recorders and photodetector modules. The photodetector modules are based on avalanche photodiodes and low-noise transimpedance amplifiers. The recording module design uses switched-capacitor array technology. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00933813
Volume :
47
Issue :
6
Database :
Academic Search Index
Journal :
IEEE Transactions on Plasma Science
Publication Type :
Academic Journal
Accession number :
137270890
Full Text :
https://doi.org/10.1109/TPS.2019.2910795