Back to Search Start Over

A HEMT-based cryogenic charge amplifier with sub-100 eVee ionization resolution for massive semiconductor dark matter detectors.

Authors :
Phipps, A.
Juillard, A.
Sadoulet, B.
Serfass, B.
Jin, Y.
Source :
Nuclear Instruments & Methods in Physics Research Section A. Oct2019, Vol. 940, p181-184. 4p.
Publication Year :
2019

Abstract

We present the measured baseline ionization resolution of a HEMT-based cryogenic charge amplifier coupled to a CDMS-II detector. The amplifier has been developed to allow massive semiconductor dark matter detectors to retain background discrimination at the low recoil energies produced by low-mass WIMPs. We find a calibrated baseline ionization resolution of σ E = 91 eV e e. To our knowledge, this is the best direct ionization resolution achieved with such massive (≈ 150 pF capacitance) radiation detectors. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
01689002
Volume :
940
Database :
Academic Search Index
Journal :
Nuclear Instruments & Methods in Physics Research Section A
Publication Type :
Academic Journal
Accession number :
137432002
Full Text :
https://doi.org/10.1016/j.nima.2019.06.022