Back to Search
Start Over
A HEMT-based cryogenic charge amplifier with sub-100 eVee ionization resolution for massive semiconductor dark matter detectors.
- Source :
-
Nuclear Instruments & Methods in Physics Research Section A . Oct2019, Vol. 940, p181-184. 4p. - Publication Year :
- 2019
-
Abstract
- We present the measured baseline ionization resolution of a HEMT-based cryogenic charge amplifier coupled to a CDMS-II detector. The amplifier has been developed to allow massive semiconductor dark matter detectors to retain background discrimination at the low recoil energies produced by low-mass WIMPs. We find a calibrated baseline ionization resolution of σ E = 91 eV e e. To our knowledge, this is the best direct ionization resolution achieved with such massive (≈ 150 pF capacitance) radiation detectors. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 01689002
- Volume :
- 940
- Database :
- Academic Search Index
- Journal :
- Nuclear Instruments & Methods in Physics Research Section A
- Publication Type :
- Academic Journal
- Accession number :
- 137432002
- Full Text :
- https://doi.org/10.1016/j.nima.2019.06.022