Development of a multi-segmented proportional gas counter for β-decay spectroscopy at KISS.
- Source :
-
Nuclear Instruments & Methods in Physics Research Section B . Jan2020, Vol. 463, p421-424. 4p. - Publication Year :
- 2020
-
Abstract
- We developed a multi-segmented proportional gas counter to perform β -decay spectroscopy of neutron-rich nuclei extracted from KEK Isotope Separation System (KISS). The 32 segmented gas counters allow us two-dimensional tracking of the events to identify β -ray events by discriminating background events. The detectors being by low-density material of thin aluminized Mylar foils and P10 gas enabled us to detect low energy electrons down to 100 keV. The absolute detection efficiency was 45% at Q β = 1 MeV, and the measured background event rate was 0.11 cps. The detector was applied to perform β -decay spectroscopy and hyperfine structure measurements of neutron-rich isotopes of Pt, Ir, and Os. We started to develop the gas counter with three-dimensional tracking by replacing the material of anode wire from BeCu to resistive thin carbon to reduce the background event rate below 0.01 cps. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 0168583X
- Volume :
- 463
- Database :
- Academic Search Index
- Journal :
- Nuclear Instruments & Methods in Physics Research Section B
- Publication Type :
- Academic Journal
- Accession number :
- 140876177
- Full Text :
- https://doi.org/10.1016/j.nimb.2019.04.036