Breath Analysis: Comparison among Methodological Approaches for Breath Sampling.

Authors :
Di Gilio, Alessia
Palmisani, Jolanda
Ventrella, Gianrocco
Facchini, Laura
Catino, Annamaria
Varesano, Niccolò
Pizzutilo, Pamela
Galetta, Domenico
Borelli, Massimo
Barbieri, Pierluigi
Licen, Sabina
de Gennaro, Gianluigi
Perbellini, Luigi
Monasta, Lorenzo
Tsakal, Andreas
Source :
Molecules. Dec2020, Vol. 25 Issue 24, p5823. 1p.
Publication Year :
2020

Abstract

Despite promising results obtained in the early diagnosis of several pathologies, breath analysis still remains an unused technique in clinical practice due to the lack of breath sampling standardized procedures able to guarantee a good repeatability and comparability of results. The most diffuse on an international scale breath sampling method uses polymeric bags, but, recently, devices named Mistral and ReCIVA, able to directly concentrate volatile organic compounds (VOCs) onto sorbent tubes, have been developed and launched on the market. In order to explore performances of these new automatic devices with respect to sampling in the polymeric bag and to study the differences in VOCs profile when whole or alveolar breath is collected and when pulmonary wash out with clean air is done, a tailored experimental design was developed. Three different breath sampling approaches were compared: (a) whole breath sampling by means of Tedlar bags, (b) the end-tidal breath collection using the Mistral sampler, and (c) the simultaneous collection of the whole and alveolar breath by using the ReCIVA. The obtained results showed that alveolar fraction of breath was relatively less affected by ambient air (AA) contaminants (p-values equal to 0.04 for Mistral and 0.002 for ReCIVA Low) with respect to whole breath (p-values equal to 0.97 for ReCIVA Whole). Compared to Tedlar bags, coherent results were obtained by using Mistral while lower VOCs levels were detected for samples (both breath and AA) collected by ReCIVA, likely due to uncorrected and fluctuating flow rates applied by this device. Finally, the analysis of all data also including data obtained by explorative analysis of the unique lung cancer (LC) breath sample showed that a clean air supply might determine a further confounding factor in breath analysis considering that lung wash-out is species-dependent. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
14203049
Volume :
25
Issue :
24
Database :
Academic Search Index
Journal :
Molecules
Publication Type :
Academic Journal
Accession number :
147778662
Full Text :
https://doi.org/10.3390/molecules25245823