Range probing as a quality control tool for CBCT‐based synthetic CTs: In vivo application for head and neck cancer patients.

Authors :
Seller Oria, Carmen
Thummerer, Adrian
Free, Jeffrey
Langendijk, Johannes A.
Both, Stefan
Knopf, Antje C.
Meijers, Arturs
Source :
Medical Physics. Aug2021, Vol. 48 Issue 8, p4498-4505. 8p.
Publication Year :
2021

Abstract

Purpose: Cone‐beam CT (CBCT)‐based synthetic CTs (sCT) produced with a deep convolutional neural network (DCNN) show high image quality, suggesting their potential usability in adaptive proton therapy workflows. However, the nature of such workflows involving DCNNs prevents the user from having direct control over their output. Therefore, quality control (QC) tools that monitor the sCTs and detect failures or outliers in the generated images are needed. This work evaluates the potential of using a range‐probing (RP)‐based QC tool to verify sCTs generated by a DCNN. Such a RP QC tool experimentally assesses the CT number accuracy in sCTs. Methods: A RP QC dataset consisting of repeat CTs (rCT), CBCTs, and RP acquisitions of seven head and neck cancer patients was retrospectively assessed. CBCT‐based sCTs were generated using a DCNN. The CT number accuracy in the sCTs was evaluated by computing relative range errors between measured RP fields and RP field simulations based on rCT and sCT images. Results: Mean relative range errors showed agreement between measured and simulated RP fields, ranging from −1.2% to 1.5% in rCTs, and from −0.7% to 2.7% in sCTs. Conclusions: The agreement between measured and simulated RP fields suggests the suitability of sCTs for proton dose calculations. This outcome brings sCTs generated by DCNNs closer toward clinical implementation within adaptive proton therapy treatment workflows. The proposed RP QC tool allows for CT number accuracy assessment in sCTs and can provide means of in vivo range verification. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00942405
Volume :
48
Issue :
8
Database :
Academic Search Index
Journal :
Medical Physics
Publication Type :
Academic Journal
Accession number :
151958398
Full Text :
https://doi.org/10.1002/mp.15020