Back to Search Start Over

Moisture Determination in Single Peanut Pods by Complex RF Impedance Measurement.

Authors :
Kandala, C. V. K.
Source :
IEEE Transactions on Instrumentation & Measurement. Dec2004, Vol. 53 Issue 6, p1493-1496. 4p.
Publication Year :
2004

Abstract

An assessment was made of estimating the moisture content (mc) of single peanut pods (in-shell peanuts) by measuring the complex impedance of a parallel-plate capacitor with a single pod between the plates. Capacitance (C), phase angle (θ) and/or dissipation factor (D) were measured at 1.0 and 5.0 MHz, and the measured values at these two frequencies were used in a derived empirical equation to calculate the mc. Previous research had determined that the mc of single and small samples of shelled peanuts (kernels) could be determined by similar measurements at 1 and 4.5 MHz. In this paper, the mc of the pods tested was between 5% and 25%. The predicted mc of the pods was within 1% of the standard air-oven values for about 85% of the single pods. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00189456
Volume :
53
Issue :
6
Database :
Academic Search Index
Journal :
IEEE Transactions on Instrumentation & Measurement
Publication Type :
Academic Journal
Accession number :
15320452
Full Text :
https://doi.org/10.1109/TIM.2004.834058