Back to Search Start Over

PandaX-4T cryogenic distillation system for removing krypton from xenon.

Authors :
Yan, Rui
Wang, Zhou
Cui, Xiangyi
Ju, Yonglin
Sha, Haidong
Li, Shuaijie
Huang, Peiyao
Wang, Xiuli
Ma, Wenbo
Fan, Yingjie
Zhao, Li
Liu, Jianglai
Ji, Xiangdong
Zhou, Jifang
Shang, Changsong
Liu, Liqiang
Source :
Review of Scientific Instruments. Dec2021, Vol. 92 Issue 12, p1-9. 9p.
Publication Year :
2021

Abstract

An efficient cryogenic distillation system was designed and constructed for the PandaX-4T dark matter detector based on the McCabe–Thiele method and the conservation of mass and energy. This distillation system is designed to reduce the concentration of krypton in commercial xenon from 5 × 10−7 to ∼ 1 0 − 14 mol/mol with 99% xenon collection efficiency at a maximum flow rate of 10 kg/h. The offline distillation operation has been completed and 5.75 tons of ultra-high purity xenon was produced, which is used as the detection medium in the PandaX-4T detector. The krypton concentration of the product xenon is measured with an upper limit of 8.0 ppt. The construction, operation, and stable purification performance of the cryogenic distillation system are studied with the experimental data, which is important for theoretical research and distillation operation optimization. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00346748
Volume :
92
Issue :
12
Database :
Academic Search Index
Journal :
Review of Scientific Instruments
Publication Type :
Academic Journal
Accession number :
154429614
Full Text :
https://doi.org/10.1063/5.0065154