Back to Search Start Over

Catheter knot around the mitral valve apparatus: An exceptional complication of remote magnetic navigation.

Authors :
Benali, Karim
Hammache, Nefissa
Sellal, Jean‐Marc
Mazen, Elfarra
de Chillou, Christian
Source :
Journal of Cardiovascular Electrophysiology. Jan2022, Vol. 33 Issue 1, p137-139. 3p. 1 Color Photograph, 1 Diagram, 1 Graph.
Publication Year :
2022

Abstract

Remote magnetic navigation (RMN) is as safe and effective as manual navigation for catheter ablation of ventricular arrhythmias. This case is the first description of a soft‐tip ablation catheter entrapment in the mitral valve apparatus during an RMN ablation procedure. The tight knot created by the catheter around a mitral valve chordae required surgical removal. This complication, which has never been reported before, highlights the need for closer fluoroscopic monitoring when performing catheter loops inside the ventricles when using the RMN system. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
10453873
Volume :
33
Issue :
1
Database :
Academic Search Index
Journal :
Journal of Cardiovascular Electrophysiology
Publication Type :
Academic Journal
Accession number :
154565664
Full Text :
https://doi.org/10.1111/jce.15313