Back to Search Start Over

Twelve-month clinical outcomes after implantation of a new, modular, anterior shape-changing fluid optic intraocular lens.

Authors :
Garg, Sumit
De Jesus, Melinda T.
Fletcher, Laura M.
Chayet, Arturo
Barragán, Enrique
Casey, Patrick
Source :
Journal of Cataract & Refractive Surgery. Oct2022, Vol. 48 Issue 10, p1134-1140. 7p.
Publication Year :
2022

Abstract

This exploratory study of a new anterior shape-changing presbyopia-correcting intraocular lens demonstrated good outcomes for distance, intermediate, and near vision, low rates of dysphotopsia, and an acceptable safety profile. Purpose: To report the initial safety and effectiveness profile for an anterior shape-changing, modular IOL, Juvene IOL (JIOL), for the treatment of aphakia and presbyopia after removal of the natural crystalline lens due to cataract. Setting: 2 private practices in Monterrey and Tijuana, Mexico. Design: Exploratory prospective multicenter open-label noncomparative clinical trial. Methods: A convenience sample of patients aged 50 to 80 years with planned cataract surgery was recruited to undergo unilateral or bilateral implantation with the JIOL. Patients were required to complete an informed consent and be able to dilate to at least 6.0 mm pharmacologically, be in good overall health, and have no significant eye health history to qualify. Visual acuities, defocus curves, and contrast sensitivity were measured for all patients 12 months postoperatively. Results: 51 of 58 eyes completed the 12-month visit. Intraoperative complication rates were extremely low (n = 1; missed base lens tab). The most frequent adverse events (AEs) were prolonged inflammation (N = 6) and cystoid macular edema (N = 4); all AEs were resolved without sequelae by the 12-month visit. The mean monocular logMAR corrected distance visual acuity, distance-corrected intermediate visual acuity, and distance-corrected near visual acuity were 0.01, 0.08, and 0.24, respectively. Defocus testing showed visual acuity > 20/40 from approximately +1.00 through −2.00 diopters. Binocular implantation (n = 16) provided superior performance over monocular implantation. Conclusions: The JIOL offers a new solution to treat presbyopia, providing clear functional vision performance across a range of distances with an acceptable initial safety profile. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
08863350
Volume :
48
Issue :
10
Database :
Academic Search Index
Journal :
Journal of Cataract & Refractive Surgery
Publication Type :
Academic Journal
Accession number :
159443072
Full Text :
https://doi.org/10.1097/j.jcrs.0000000000000935