Back to Search Start Over

A Highly Sensitive Fluorescence and Screen-Printed Electrodes—Electrochemiluminescence Immunosensor for Ricin Detection Based on CdSe/ZnS QDs with Dual Signal.

Authors :
Feng, Shasha
Hu, Wei
Pei, Fubin
Liu, Zhiwei
Du, Bin
Mu, Xihui
Liu, Bing
Hao, Qingli
Lei, Wu
Tong, Zhaoyang
Source :
Toxins. Oct2022, Vol. 14 Issue 10, pN.PAG-N.PAG. 12p.
Publication Year :
2022

Abstract

A sensitive dual-readout immunosensor for fluorescence and electrochemiluminescence (ECL) detection of ricin was established, which was combined with a streptavidin–biotin signal amplification system. CdSe/ZnS quantum dots with fine fluorescence and ECL properties were used as the dual-signal function probes of the sandwich immunocomplex. Under the optimum experimental conditions, the dual signal intensity increased significantly with the rise in ricin concentration. The fluorescence intensity of the senor exhibited a good liner relationship toward the ricin concentrations with 0.1~100 ng/mL and the limit of detection (LOD) was 81.7 pg/mL; taking ECL as the detection signal, the sensor showed a linear relationship with the ricin concentrations ranging from 0.01 ng/mL to 100 ng/mL and the LOD was 5.5 pg/mL. The constructed sensor with high sensitivity had been successfully applied to the detection of ricin in complex matrices with satisfactory recoveries. The proposed immunosensor model can be extended to the analysis and detection of others target proteins. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
20726651
Volume :
14
Issue :
10
Database :
Academic Search Index
Journal :
Toxins
Publication Type :
Academic Journal
Accession number :
159962525
Full Text :
https://doi.org/10.3390/toxins14100710