Back to Search Start Over

Design of a S-shape middle-ear ossicular replacement prosthesis and its comparison with present-day prostheses using finite element modelling.

Authors :
Seivur, Sundaram
Rathnakara, Subhodha H
Ananthasuresh, G K
Source :
Sādhanā: Academy Proceedings in Engineering Sciences. Dec2022, Vol. 47 Issue 4, p1-16. 16p.
Publication Year :
2022

Abstract

The impedance-matching mechanism of the ossicular chain of the middle ear is disrupted by erosion of the incus, warranting prosthetic intervention. The present-day practice of straight rigid-rod is not competent to restore the required level of natural motion attenuation and force amplification between the tympanic membrane and stapes-footplate. In this work, we use finite element simulation of anatomically accurate middle ear to evaluate the efficacy. We compare the performance of differently shaped incus-prosthesis, T, C, and S in particular, using simulation. We find that the S-shaped incus-prosthesis gives the best performance. We also prototyped the prosthesis and performed a preliminary cadaveric study. The finite element simulation combined with the ability to 3D-print has the potential for effective personalized surgical intervention. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
02562499
Volume :
47
Issue :
4
Database :
Academic Search Index
Journal :
Sādhanā: Academy Proceedings in Engineering Sciences
Publication Type :
Academic Journal
Accession number :
160192932
Full Text :
https://doi.org/10.1007/s12046-022-02000-3