Back to Search
Start Over
A position-sensitive large-area microchannel plate detector with digital data-acquisition system for studies of exotic nuclei.
- Source :
-
Nuclear Instruments & Methods in Physics Research Section B . Aug2023, Vol. 541, p232-235. 4p. - Publication Year :
- 2023
-
Abstract
- Development of a position-sensitive large-area microchannel plate (MCP) detector based detection system is reported. Two MCPs with 120 mm active diameter, mounted in chevron configuration, with delay-line anode are read out by a digital data-acquisition system consists of fast-timing amplifier unit and CAEN v1751 waveform digitizer. Details of the digital pulse processing, programmed triggering and calibration setup are described. The obtained position resolution was studied using α -particles, the best resolutions on X and Y axes are 460 ± 1 μ m and 579 ± 1 μ m in FWHM, respectively. [ABSTRACT FROM AUTHOR]
- Subjects :
- *MICROCHANNEL plates
*EXOTIC nuclei
*DETECTORS
*COINCIDENCE
*ANODES
Subjects
Details
- Language :
- English
- ISSN :
- 0168583X
- Volume :
- 541
- Database :
- Academic Search Index
- Journal :
- Nuclear Instruments & Methods in Physics Research Section B
- Publication Type :
- Academic Journal
- Accession number :
- 164250638
- Full Text :
- https://doi.org/10.1016/j.nimb.2023.05.032