Back to Search Start Over

Applicability of the Meniscus-Removal Method for Q-Band Liquid Characterization in Semi-Open Waveguide Cell †.

Authors :
Kalisiak, Michał
Wiatr, Wojciech
Papis, Radosław
Source :
Sensors (14248220). Jun2023, Vol. 23 Issue 12, p5390. 12p.
Publication Year :
2023

Abstract

We present the broadband transmission-reflection meniscus-removal method for liquid characterization in a semi-open rectangular waveguide. The algorithm utilizes 2-port scattering parameters measured with a calibrated vector network analyzer for three states of the measurement cell: empty and filled with two liquid levels. The method enables the mathematical de-embedding of a symmetrical sample of a liquid, not distorted with a meniscus, and provision of its permittivity and permeability, as well as its height. We validate the method for propan-2-ol (IPA), a 50% aqueous solution of IPA, and distilled water in the Q-band (33–50 GHz). We investigate typical problems for in-waveguide measurements, such as phase ambiguity. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
14248220
Volume :
23
Issue :
12
Database :
Academic Search Index
Journal :
Sensors (14248220)
Publication Type :
Academic Journal
Accession number :
164724226
Full Text :
https://doi.org/10.3390/s23125390