Experiences with continuous intra-arterial blood gas monitoring: precision and drift of a pure optode-system.

Authors :
Menzel, Matthias
Soukup, Jens
Henze, Dirk
Engelbrecht, Kai
Senderreck, Maik
Scharf, Alexander
Rieger, Andreas
Grond, Stefan
Source :
Intensive Care Medicine. Dec2003, Vol. 29 Issue 12, p2180-2186. 7p.
Publication Year :
2003

Abstract

<bold>Objective: </bold>The utility of continuous intra-arterial blood gas analysis (CBGA) with combined electrochemical and optode sensors has been demonstrated. More recently, a pure optode sensor with a changed sensing element architecture has become available. The aim was to determine the measurement accuracy and long-term stability of the new sensor.<bold>Design: </bold>A prospective explorative study was performed. Simultaneous measurements of intermittent blood gas analyses (IBGA) (ABL 610, Radiometer, Copenhagen) and CBGA (Diametrics Medical, High Wycombe, Bucks., UK) were compared using Bland-Altman analysis.<bold>Patients: </bold>Twenty-five patients admitted to the ICU and requiring mechanical ventilation for an expected minimum of about 96 h were included.<bold>Results: </bold>Mean monitoring time was 106.1 (range 15-231) hours. Bias and precision for PO(2 )were -0.2 kPa (1%)+/-1.8 kPa (9.5%); PCO(2): 0.03 kPa (0.6%)+/-0.44 kPa (9.3%); pH: -0.001 (0.01%)+/-0.04 (0.45%). The sensor showed no change of measurement characteristics during 4 days of measurement. However, in 69 cases continuous monitoring was interrupted (reversible sudden drops of PO(2) measurement) possibly caused by thrombotic deposition and/or sensor bending and accidental sensor retraction.<bold>Conclusions: </bold>The precision and bias of the PCO(2)- and pH-sensing elements were in line with the findings of the older sensor technology. The possibility that the PO(2) optode could offer greater accuracy than the older technology is suggested by comparisons with results reported in previous studies. No sensor drift occurred during long-term measurement over more than 4 days. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
03424642
Volume :
29
Issue :
12
Database :
Academic Search Index
Journal :
Intensive Care Medicine
Publication Type :
Academic Journal
Accession number :
16629183
Full Text :
https://doi.org/10.1007/s00134-003-1962-1