Back to Search Start Over

Severe pulmonary valve insufficiency caused by transjugular cannulation of pulmonary artery for right ventricular assist device: diagnosis and surgical solution—a case report.

Authors :
Wert, Leonhard
Lanmüller, Pia
Ott, Sascha
Mulzer, Johanna
Starck, Christoph T
Falk, Volkmar
Potapov, Evgenij V
Source :
Journal of Surgical Case Reports. Jul2023, Vol. 2023 Issue 7, p1-3. 3p.
Publication Year :
2023

Abstract

Implantation of a temporary percutaneous right ventricular assist device (RVAD) in patients with right heart failure after left ventricular assist device (LVAD) implantation is an established technique that may cause complications. We present a 60-year-old male patient who underwent urgent LVAD implantation. On the second postoperative day the patient developed acute right heart failure. We implanted a temporary percutaneous RVAD with two cannulas via the right internal jugular vein and the right femoral vein. Transesophageal echocardiography revealed severe pulmonary insufficiency. After performing re-sternotomy we anastomosed a prosthetic graft to the pulmonary trunk (PT), performed subxiphoid tunneling of the graft and replaced the transjugular outflow cannula. The pulmonary regurgitation caused by the percutaneous transvalvular cannula disappeared. In such case a direct anastomosis to the PT is the solution. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
20428812
Volume :
2023
Issue :
7
Database :
Academic Search Index
Journal :
Journal of Surgical Case Reports
Publication Type :
Academic Journal
Accession number :
169851020
Full Text :
https://doi.org/10.1093/jscr/rjad389