Back to Search Start Over

Intermacs quarterly report analysis to monitor longitudinal outcomes in a centrifugal flow assist device.

Authors :
Li, Song
Slaughter, Mark S.
Hall, Shelley
Davis, Erin
Kappetein, Arie Pieter
Dedrick, Alexandra
Salerno, Christopher T.
Pagani, Francis D.
Mahr, Claudius
Source :
Journal of Heart & Lung Transplantation. Nov2023, Vol. 42 Issue 11, p1510-1514. 5p.
Publication Year :
2023

Abstract

In June 2021, HVAD System distribution ceased due to observational data demonstrating increased mortality and neurological events compared to another commercial device, and a device malfunction with delay or failure to restart, especially in certain subpopulations. To assess ongoing risk for patients on support following subsequent device recalls, the manufacturer's Intermacs HVAD System 2022 Quarterly Reports were queried to identify mortality and adverse events trends in a contemporary cohort of 3110 primary HVAD implantations since October 2017, stratified by year-of-implant. Mean duration of support was 21 ± 16 months, with 33% alive on original device, 25% transplanted, 6% undergoing device exchange, 4% recovered, and 32% expired. Kaplan–Meier and event-per-patient-year estimates for survival, freedom from device explant, stroke, and pump thrombus were similar across year-of-implant. Following market withdrawal and recent device recalls, there appears to be no increase in mortality, stroke, pump thrombus, or explant for HVAD-supported patients. Quarterly report monitoring is ongoing. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
10532498
Volume :
42
Issue :
11
Database :
Academic Search Index
Journal :
Journal of Heart & Lung Transplantation
Publication Type :
Academic Journal
Accession number :
172978604
Full Text :
https://doi.org/10.1016/j.healun.2023.07.008