Back to Search
Start Over
Standardization of rapid screening technique based on liquid scintillation counting for estimation of gross α/β activities in urine.
- Source :
-
Journal of Radioanalytical & Nuclear Chemistry . Jan2024, Vol. 333 Issue 1, p179-192. 14p. - Publication Year :
- 2024
-
Abstract
- As a part of radiation emergency response, a rapid method for estimation of gross-α and gross-β activities in urine samples using liquid scintillation counting technique was standardized and validated using spiked urine samples. Effect of various storage conditions such as temperature, storage time and use of preservatives on estimation of gross-α and gross-β activities was investigated systematically. Measurement results revealed that urine samples either should be stored at lower temperatures (≤ 5 °C) or should be acidified with 1% v/v HCl for quantification of gross-α and gross-β activities with bias ≤ 10%. [ABSTRACT FROM AUTHOR]
- Subjects :
- *LIQUID scintillation counting
*URINE
*SCINTILLATORS
*STANDARDIZATION
Subjects
Details
- Language :
- English
- ISSN :
- 02365731
- Volume :
- 333
- Issue :
- 1
- Database :
- Academic Search Index
- Journal :
- Journal of Radioanalytical & Nuclear Chemistry
- Publication Type :
- Academic Journal
- Accession number :
- 174879368
- Full Text :
- https://doi.org/10.1007/s10967-023-09277-6