Back to Search Start Over

Practical application of the patient data-based quality control method: the potassium example.

Authors :
Yan Zhang
Hua-Li Wang
Ye-Hong Xie
Da-Hai He
Chao-Qiong Zhou
Li-Rui Kong
Source :
Biochemia Medica. 2024, Vol. 34 Issue 1, p1-5. 5p.
Publication Year :
2024

Abstract

Introduction: Internal quality control (IQC) is a core pillar of laboratory quality control strategies. Internal quality control commercial materials lack the same characteristics as patient samples and IQC contributes to the costs of laboratory testing. Patient data-based quality control (PDB-QC) may be a valuable supplement to IQC; the smaller the biological variation, the stronger the ability to detect errors. Using the potassium concentration in serum as an example study compared error detection effectiveness between PDB-QC and IQC. Materials and methods: Serum potassium concentrations were measured by using an indirect ion-selective electrode method. For the training database, 23,772 patient-generated data and 366 IQC data from April 2022 to September 2022 were used; 15,351 patient-generated data and 246 IQC data from October 2022 to January 2023 were used as the testing database. For both PDB-QC and IQC, average values and standard deviations were calculated, and z-score charts were plotted for comparison purposes. Results: Five systematic and three random errors were detected using IQC. Nine systematic errors but no random errors were detected in PDB-QC. The PDB-QC showed systematic error warnings earlier than the IQC. Conclusions: The daily average value of patient-generated data was superior to IQC in terms of the efficiency and timeliness of detecting systematic errors but inferior to IQC in detecting random errors. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
13300962
Volume :
34
Issue :
1
Database :
Academic Search Index
Journal :
Biochemia Medica
Publication Type :
Academic Journal
Accession number :
175734818
Full Text :
https://doi.org/10.11613/BM.2024.010901