T‐wave oversensing after implantation of pacemakers with automatic sensitivity adjustment in patients with atrioventricular block.

Authors :
Takahashi, Masao
Kujiraoka, Hirofumi
Arai, Tomoyuki
Hojo, Rintaro
Fukamizu, Seiji
Source :
Pacing & Clinical Electrophysiology. Apr2024, Vol. 47 Issue 4, p561-563. 3p.
Publication Year :
2024

Abstract

T‐wave oversensing in pacemakers is rare because the set sensitivity is generally fixed. However, several models of pacemaker employ automatic sensitivity adjustment. Here, we present two cases of atrioventricular block treated by implantation of the pacemaker with automatic sensitivity adjustment. After implanting the pacemaker with automatic sensitivity adjustment, ventricular pacing suppression due to T‐wave oversensing occurred. In both cases, T‐wave oversensing disappeared after adjusting the setting sensitivity from 0.9 to 2.0 mV. [ABSTRACT FROM AUTHOR]