Illumination matters part IV: blackout and whiteout in flexible ureteroscopy - first report on a phenomenon observed by PEARLS.

Authors :
Kwok, Jia-Lun
De Coninck, Vincent
Panthier, Frédéric
Kamkoum, Hatem
Pauchard, Felipe
Shrestha, Anil
Gauhar, Vineet
Brachlow, Jan
Schmid, Florian Alexander
Poyet, Cédric
Eberli, Daniel
Traxer, Olivier
Keller, Etienne Xavier
Source :
World Journal of Urology. 2024, Vol. 42 Issue 1, p1-7. 7p.
Publication Year :
2024

Abstract

Purpose: To date, no study has evaluated effects of varying brightness settings on image quality from flexible ureteroscopes submerged in saline. The aim was to evaluate blackout and whiteout occurrences in an in-vitro kidney calyx model. Material and methods: We evaluated a series of contemporary flexible ureteroscopes including the Storz Flex-Xc and Flex-X2s, Olympus V3 and P7, Pusen 7.5F and 9.2F, as well as OTU WiScope using a 3D-printed enclosed pink in-vitro kidney calyx model submerged in saline. Endoscopic images were captured with ureteroscope tip placed at 5 mm,10 mm and 20 mm distances. The complete range of brightness settings and video capture modes were evaluated for each scope. Distribution of brightness on a grayscale histogram of images was analyzed (scale range 0 to 255). Blackout and whiteout were defined as median histogram ranges from 0 to 35 and 220 to 255, respectively (monitor image too dark or too bright for the human eye, respectively). Results: Blackout occurred with the P7, Pusen 7.5F, 9.2F and WiScope at all distances, and V3 at 20 mm - with lowest brightness settings. Whiteout occurred with Flex-X2s, V3 and P7 at 5 mm and 10 mm, as well as with V3 and P7 at 20 mm - mostly with highest brightness settings. The Flex-Xc had neither blackout nor whiteout at all settings and distances. Conclusion: Blackout or whiteout of images is an undesirable property that was found for several scopes, possibly impacting diagnostic and therapeutic purposes during ureteroscopy. These observations form a guide to impact a urologist’s choice of instruments and settings. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
07244983
Volume :
42
Issue :
1
Database :
Academic Search Index
Journal :
World Journal of Urology
Publication Type :
Academic Journal
Accession number :
177057960
Full Text :
https://doi.org/10.1007/s00345-024-04988-1