Back to Search Start Over

Commercially-available AI algorithm improves radiologists' sensitivity for wrist and hand fracture detection on X-ray, compared to a CT-based ground truth.

Authors :
Jacques, Thibaut
Cardot, Nicolas
Ventre, Jeanne
Demondion, Xavier
Cotten, Anne
Source :
European Radiology. May2024, Vol. 34 Issue 5, p2885-2894. 10p.
Publication Year :
2024

Abstract

Objectives: Algorithms for fracture detection are spreading in clinical practice, but the use of X-ray-only ground truth can induce bias in their evaluation. This study assessed radiologists' performances to detect wrist and hand fractures on radiographs, using a commercially-available algorithm, compared to a computerized tomography (CT) ground truth. Methods: Post-traumatic hand and wrist CT and concomitant X-ray examinations were retrospectively gathered. Radiographs were labeled based on CT findings. The dataset was composed of 296 consecutive cases: 118 normal (39.9%), 178 pathological (60.1%) with a total of 267 fractures visible in CT. Twenty-three radiologists with various levels of experience reviewed all radiographs without AI, then using it, blinded towards CT results. Results: Using AI improved radiologists' sensitivity (Se, 0.658 to 0.703, p < 0.0001) and negative predictive value (NPV, 0.585 to 0.618, p < 0.0001), without affecting their specificity (Sp, 0.885 vs 0.891, p = 0.91) or positive predictive value (PPV, 0.887 vs 0.899, p = 0.08). On the radiographic dataset, based on the CT ground truth, stand-alone AI performances were 0.771 (Se), 0.898 (Sp), 0.684 (NPV), 0.915 (PPV), and 0.764 (AUROC) which were lower than previously reported, suggesting a potential underestimation of the number of missed fractures in the AI literature. Conclusions: AI enabled radiologists to improve their sensitivity and negative predictive value for wrist and hand fracture detection on radiographs, without affecting their specificity or positive predictive value, compared to a CT-based ground truth. Using CT as gold standard for X-ray labels is innovative, leading to algorithm performance poorer than reported elsewhere, but probably closer to clinical reality. Clinical relevance statement: Using an AI algorithm significantly improved radiologists' sensitivity and negative predictive value in detecting wrist and hand fractures on radiographs, with ground truth labels based on CT findings. Key Points: • Using CT as a ground truth for labeling X-rays is new in AI literature, and led to algorithm performance significantly poorer than reported elsewhere (AUROC: 0.764), but probably closer to clinical reality. • AI enabled radiologists to significantly improve their sensitivity (+ 4.5%) and negative predictive value (+ 3.3%) for the detection of wrist and hand fractures on X-rays. • There was no significant change in terms of specificity or positive predictive value. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
09387994
Volume :
34
Issue :
5
Database :
Academic Search Index
Journal :
European Radiology
Publication Type :
Academic Journal
Accession number :
177463593