Development of a quantum cascade laser absorption spectrometer for simultaneous measurement of 13C–18O and 18O–18O clumping in CO2.

Authors :
Wieman, Scott T.
Kapit, Jason
Michel, Anna P. M.
Guo, Weifu
Source :
Rapid Communications in Mass Spectrometry: RCM. 8/30/2024, Vol. 38 Issue 16, p1-11. 11p.
Publication Year :
2024

Abstract

Rationale: Dual clumped isotope paleothermometry determines carbonate formation temperatures by measuring the frequency of 13C–18O (∆638) and 18O–18O (∆828) pairs in carbonates. It resolves isotopic kinetic biases and thus enables more accurate paleotemperature reconstructions. However, high‐precision measurements of 18O–18O clumping using current techniques requires large sample sizes and long acquisition times. Methods: We developed a mid‐infrared isotope ratio laser spectrometer (IRLS) for simultaneous measurement of the isotopologue ratios ∆638 and ∆828 in gas‐phase carbon dioxide (CO2) at room temperature. Our IRLS uses a single laser scanning from 2290.7 to 2291.1 cm−1 and a 31 m pathlength optical cell, and it simultaneously measures the five isotopologues required for calculating ∆638 and ∆828: 16O12C16O, 16O13C16O, 16O12C18O, 16O13C18O, and 18O12C18O. In addition, our IRLS can measure 16O12C17O, enabling ∆17O analysis. Results: At ~20°C and a CO2 pressure of ~2 Torr, our IRLS system achieved precisions of 0.128‰ and 0.140‰ within 20 s for abundances of the clumped isotopologues 16O13C18O and 18O12C18O, respectively, and precisions of 0.267‰, 0.245‰, and 0.128‰ for 16O12C16O, 16O13C16O, and 16O12C18O. This yielded precisions of 0.348‰ (∆638) and 0.302‰ (∆828) within 25 s. Simulated sample–reference switching highlights the potential of our system and the need for further development. Conclusions: We demonstrated simultaneous measurements of ∆638 and ∆828 in CO2 to precisions of <0.35‰ within 25 s using a room‐temperature, single‐laser IRLS. Future developments on better resolving 16O12C16O and 16O13C16O peaks and system temperature control could further improve the measurement precision. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
09514198
Volume :
38
Issue :
16
Database :
Academic Search Index
Journal :
Rapid Communications in Mass Spectrometry: RCM
Publication Type :
Academic Journal
Accession number :
178426634
Full Text :
https://doi.org/10.1002/rcm.9836