Back to Search
Start Over
The accuracy of radial artery applanation tonometry and intra‐arterial blood pressure monitoring in critically ill patients: An evidence‐based review.
- Source :
-
Nursing in Critical Care . Nov2024, Vol. 29 Issue 6, p1751-1757. 7p. - Publication Year :
- 2024
-
Abstract
- The invasive intra‐arterial approach is the gold standard for measuring blood pressure in intensive care units where accuracy is crucial. However, invasive procedures increase the risk of infections and mortality. This evidence‐based review aimed to determine whether continuous non‐invasive blood pressure (CNIBP) monitoring, using Radial Artery Applanation Tonometry (RAAT) devices, is as accurate as invasive methods. Six papers were included: three prospective cohort studies and three comparative studies. Most studies showed that mean arterial pressure is accurately recorded through RAAT monitoring; however, more research is needed to assess the accuracy of non‐invasive readings of systolic and diastolic blood pressures, as data are not always concordant. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 13621017
- Volume :
- 29
- Issue :
- 6
- Database :
- Academic Search Index
- Journal :
- Nursing in Critical Care
- Publication Type :
- Academic Journal
- Accession number :
- 180562031
- Full Text :
- https://doi.org/10.1111/nicc.13006