Back to Search
Start Over
Extensive Deuterium Back-Exchange in Certain Immobilized Pepsin Columns Used for H/D Exchange Mass Spectrometry.
- Source :
-
Analytical Chemistry . 3/1/2006, Vol. 78 Issue 5, p1719-1723. 5p. 7 Diagrams. - Publication Year :
- 2006
-
Abstract
- Pepsin digestion prior to mass analysis increases the spatial resolution of hydrogen exchange mass spectrometry experiments. Online digestion with immobilized pepsin is advantageous for several reasons including better digestion efficiency. We have found that certain immobilized pepsin columns cause substantial deuterium back-exchange, rendering the data unusable. When pepsin immobilized on a POROS support was used for online digestion, back-exchange was within the expected range and was similar to the back-exchange of deuterated peptides produced by in-solution pepsin digestion. However, when pepsin immobilized onto selected polystyrene-divinylbenzene supports was used for online digestion with the same system, deuterium loss was extremely high. The effect seems linked to the properties of the solid support used to conjugate the pepsin. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 00032700
- Volume :
- 78
- Issue :
- 5
- Database :
- Academic Search Index
- Journal :
- Analytical Chemistry
- Publication Type :
- Academic Journal
- Accession number :
- 20330648
- Full Text :
- https://doi.org/10.1021/ac0518497