Back to Search Start Over

Modelling in the study of interaction of Hemopump device and artificial ventilation

Authors :
De Lazzari, C.
Darowski, M.
Ferrari, G.
Pisanelli, D.M.
Tosti, G.
Source :
Computers in Biology & Medicine. Nov2006, Vol. 36 Issue 11, p1235-1251. 17p.
Publication Year :
2006

Abstract

Abstract: The aim of this work is to evaluate in different ventricular conditions the influence of joint mechanical ventilation (MV) and Hemopump assistance. To perform this study, we used a computer simulator of human cardiovascular system where the influence of MV was introduced changing thoracic pressure to positive values. The simulation confirmed that haemodynamic variables are highly sensitive to thoracic pressure changes. On the other hand, Hemopump assistance raises, among the others, mean aortic pressure, total cardiac output (left ventricular output flow plus Hemopump flow) and coronary flow. The simulation showed that the joint action of Hemopump and positive thoracic pressure diminishes these effects. [Copyright &y& Elsevier]

Details

Language :
English
ISSN :
00104825
Volume :
36
Issue :
11
Database :
Academic Search Index
Journal :
Computers in Biology & Medicine
Publication Type :
Academic Journal
Accession number :
22134318
Full Text :
https://doi.org/10.1016/j.compbiomed.2005.08.001