Back to Search
Start Over
International standards for minimum potency of anti-A and anti-B blood grouping reagents: evaluation of candidate preparations in an international collaborative study.
- Source :
-
Vox Sanguinis . Nov2006, Vol. 91 Issue 4, p336-344. 9p. 6 Diagrams. - Publication Year :
- 2006
-
Abstract
- Background and Objectives The aim of the study was to evaluate lyophilized monoclonal IgM anti-A and anti-B preparations for use as international standards (IS) to specify recommended minimum potencies of anti-A and anti-B blood grouping reagents in tube tests. Materials and Methods The candidate IS for minimum potency of anti-A and anti-B blood grouping reagents, codes 03/188 and 03/164, respectively, were evaluated against a wide range of commercial anti-A and anti-B blood grouping reagents in an international collaborative study involving 16 laboratories in nine countries. Laboratories titrated 03/188 and 03/164 in parallel with as many commercial anti-A and anti-B blood grouping reagents, respectively, as were available to them, in tube tests according to specified haemagglutination methodology. Three of these laboratories and a further laboratory also titrated 03/188 and 03/164 in parallel with currently available reference preparations for anti-A and anti-B. The ratios of the mean endpoint titres of the anti-A and anti-B reagents to those of 03/188 and 03/164, respectively, within each laboratory were calculated. Results The ratios of the mean titres of the anti-A reagents to the mean titre of 03/188 within a laboratory fell within 0·062 and 4, i.e. the potencies of the anti-A reagents were between a sixteenth to four times as strong as 03/188. The ratios of the mean titres of the anti-B reagents to the mean titre of 03/164 within a laboratory also fell within 0·062 and 4, with one outlier. Conclusions By international consensus, a 1 in 8 dilution of the candidate IS for anti-A, 03/188, and a 1 in 4 dilution of the candidate IS for anti-B, 03/164, were considered appropriate to define the recommended minimum potencies of anti-A and anti-B blood grouping reagents, respectively, in tube tests. On the basis of these results, preparations 03/188 and 03/164 were established by the World Health Organization as International Standards for Minimum Potency of Anti-A and Anti-B Blood Grouping Reagents respectively, and by the US Food and Drug Administration Center for Biologics Evaluation and Research as Minimum Potency Reference Reagents. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 00429007
- Volume :
- 91
- Issue :
- 4
- Database :
- Academic Search Index
- Journal :
- Vox Sanguinis
- Publication Type :
- Academic Journal
- Accession number :
- 23017437
- Full Text :
- https://doi.org/10.1111/j.1423-0410.2006.00834.x