Back to Search
Start Over
A Fluorescence Spot Test for Salicylate Determination.
- Source :
-
Analytical Letters . 2007, Vol. 40 Issue 3, p573-583. 11p. 1 Diagram, 2 Charts, 4 Graphs. - Publication Year :
- 2007
-
Abstract
- A procedure based on a spot test for salicylate is presented. The method is simple, rugged and is based on the salycilate native fluorescence measurement on the surface of a filter paper by using optical fibers for light conduction. The fluorescence was linear at concentration range of 0.01–0.4 mmol l-1. The relative standard deviation of 5% (n=10) was observed. Detection and quantification limits were 4 and 14 µmol l-1, respectively. The results obtained in pharmaceutical salicylate determinations showed that practically no statistical difference existed with the pharmacopeical method (95% confidence level). The recovery studies in biological samples spiked with salicylic acid at overdose concentration levels showed that the matrix sample did not interfere in the method. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 00032719
- Volume :
- 40
- Issue :
- 3
- Database :
- Academic Search Index
- Journal :
- Analytical Letters
- Publication Type :
- Academic Journal
- Accession number :
- 23518746
- Full Text :
- https://doi.org/10.1080/00032710600966010