Prevention of Inappropriate Shocks in ICD Recipients: A Review of 10,000 Tachycardia Episodes.
- Source :
-
Pacing & Clinical Electrophysiology . Jan2007 Supplement, Vol. 30, pS128-S133. 1p. 1 Diagram, 1 Chart, 1 Graph. - Publication Year :
- 2007
-
Abstract
- Background: The efficacy of dual-chamber ICD arrhythmia classification algorithms is crucial to prevent inappropriate shocks. We report our experience from a meta-analysis of five prospective clinical studies with inclusion phases ranging between 1997 and 2003. Methods: Dual-chamber ICD using standard dual-chamber arrhythmia classification algorithms were implanted in 802 patients (mean age = 64 ± 11 years, 88% men) in 74 medical centers. The ICD indication was secondary prevention in 95% of patients. Supraventricular tachyarrhythmias (SVT) were previously documented in 26% of patients. All spontaneous tachyarrhythmic events documented by the device memories were analyzed by a adjudicating committee. The episodes lasting >12 seconds and/or treated by the ICD were analyzed. Results: Over a mean follow-up of 302 ± 113 days, 9,690 events were reported. Mean heart rate at the time of events was 131 ± 45 bpm (100–430). Events were classified as oversensing in 1.4%, sinus tachycardia (ST) in 66%, SVT in 13%, slow (<150 bpm) ventricular tachycardia (VT) in 8.7%, and VT or ventricular fibrillation (VF) in 10.3%. The sensitivity of slow VT detection was 94%, and of VT/VF detection 99.3%. The specificity of sinus rhythm/ST/SVT recognition was 94%, positive predictive value 79.3%, and negative predictive value 99.2%. A total of 1,918 episodes were treated in 330 patients: 1,472 appropriately in 213, and 446 inappropriately in 117 (15% of the overall population) patients. Only 62 episodes were inappropriately treated by shocks in 40 patients, representing 5% of the overall population. Conclusions: In this conventional ICD population, the overall specificity of standard dual-chamber arrhythmia detection settings reached 94%. This feature allows efficient detection of fast as well as slow VT events with a very low rate of inappropriate shocks. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 01478389
- Volume :
- 30
- Database :
- Academic Search Index
- Journal :
- Pacing & Clinical Electrophysiology
- Publication Type :
- Academic Journal
- Accession number :
- 23710150
- Full Text :
- https://doi.org/10.1111/j.1540-8159.2007.00622.x