Back to Search Start Over

A tunable collimator for precision irradiation with ion microbeams

Authors :
Amorini, F.
Cardella, G.
Pappalardo, A.
Piscopo, M.
Sipala, V.
Finocchiaro, P.
Source :
Nuclear Instruments & Methods in Physics Research Section B. Jul2008, Vol. 266 Issue 14, p3325-3329. 5p.
Publication Year :
2008

Abstract

Abstract: We describe a cheap yet effective beam collimation system capable of providing square apertures from 1mm down to about 10μm. The tunable part is driven by two manual micrometers, one controlling the shape the other the size, which could be immediately replaced by high precision stepping motors should remote control be needed. The aperture was calibrated by means of a simple home-made laser diffraction system which allows a considerable precision. The collimation system was used to produce low-divergence beams of oxygen and protons, in order to test the position sensitivity of a silicon microstrip telescope detector. [Copyright &y& Elsevier]

Details

Language :
English
ISSN :
0168583X
Volume :
266
Issue :
14
Database :
Academic Search Index
Journal :
Nuclear Instruments & Methods in Physics Research Section B
Publication Type :
Academic Journal
Accession number :
32638635
Full Text :
https://doi.org/10.1016/j.nimb.2008.04.008