Development of a compact x-ray particle image velocimetry for measuring opaque flows.
- Source :
-
Review of Scientific Instruments . Mar2009, Vol. 80 Issue 3, p033706-033712. 7p. 1 Black and White Photograph, 4 Diagrams, 3 Charts, 5 Graphs. - Publication Year :
- 2009
-
Abstract
- A compact x-ray particle image velocimetry (PIV) system employing a medical x-ray tube as a light source was developed to measure quantitative velocity field information of opaque flows. The x-ray PIV system consists of a medical x-ray tube, an x-ray charge coupled device camera, a programmable shutter for a pulse-type x ray, and a synchronization device. Through performance tests, the feasibility of the developed x-ray PIV system as a flow measuring device was verified. To check the feasibility of the developed system, we tested a tube flow at two different mean velocities of 1 and 2 mm/s. The x-ray absorption of tracer particles must be quite different from that of working fluid to have a good contrast in x-ray images. All experiments were performed under atmospheric pressure condition. This system is unique and useful for investigating various opaque flows or flows inside opaque conduits. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 00346748
- Volume :
- 80
- Issue :
- 3
- Database :
- Academic Search Index
- Journal :
- Review of Scientific Instruments
- Publication Type :
- Academic Journal
- Accession number :
- 37259476
- Full Text :
- https://doi.org/10.1063/1.3103644