Back to Search Start Over

Neutron–photon discrimination and spectrum unfolding with a stilbene detector

Authors :
Mertens, C.
De Lellis, C.
Tondeur, F.
Source :
Applied Radiation & Isotopes. Apr2010, Vol. 68 Issue 4/5, p957-960. 4p.
Publication Year :
2010

Abstract

Abstract: As a first step into the development of a neutron dose monitor, a stilbene detector is used to test a procedure applicable to other organic scintillators allowing for neutron–photon discrimination. The pulses are measured by numerical acquisition and their amplitude and decay time are calculated by software. The discrimination is performed in an amplitude-decay time plot, and separate amplitude spectra are built. These spectra are unfolded MAXED, using the detector response matrices calculated with MCNPX. [Copyright &y& Elsevier]

Details

Language :
English
ISSN :
09698043
Volume :
68
Issue :
4/5
Database :
Academic Search Index
Journal :
Applied Radiation & Isotopes
Publication Type :
Academic Journal
Accession number :
48603132
Full Text :
https://doi.org/10.1016/j.apradiso.2009.10.014