Fully Wireless Implantable Cardiovascular Pressure Monitor Integrated with a Medical Stent.
- Source :
-
IEEE Transactions on Biomedical Engineering . Jun2010, Vol. 57 Issue 6, p1487-1496. 10p. - Publication Year :
- 2010
-
Abstract
- This paper presents a fully wireless cardiac pressure sensing system. Food and Drug Administration (FDA) approved medical stents are explored as radiating structures to support simultaneous transcutaneous wireless telemetry and powering. An application-specific integrated circuit (ASIC), designed and fabricated using the Texas Instruments 130-nm CMOS process, enables wireless telemetry, remote powering, voltage regulation, and processing of pressure measurements from a microelectromechanical systems (MEMS) capacitive sensor. This paper demonstrates fully wireless-pressure-sensing functionality with an external 35-dB⋅m RF powering source across a distance of 10 cm. Measurements in a regulated pressure chamber demonstrate the ability of the cardiac system to achieve pressure resolutions of 0.5 mmHg over a range of 0-50 mmHg using a channel data-rate of 42.2 kb/s. [ABSTRACT FROM AUTHOR]