Back to Search
Start Over
A preliminary study of the performance of a novel design of multi-element tissue equivalent proportional counter for neutron monitoring
- Source :
-
Radiation Measurements . Dec2010, Vol. 45 Issue 10, p1309-1312. 4p. - Publication Year :
- 2010
-
Abstract
- Abstract: In this study we investigate the dependence of the sensitivity of a TEPC upon its surface area and demonstrate that a compact multi-element tissue equivalent proportional counter (METEPC) has a counting sensitivity comparable to a commercially available 12.7 cm (5 inch) diameter spherical TEPC. The METEPC incorporates 61 cylindrical counting volumes of internal diameter of 0.5 cm and height 5 cm, machined in a single block of tissue equivalent plastic. It is the simplest design available in the multi-element geometrical configuration and is approximately nine times smaller in volume than that of a conventional 12.7 cm spherical TEPC. The neutron sensitivity of commercially available TEPCs and the METEPC simulating a 2 μm tissue site size was examined experimentally using the McMaster University 1.25 MV double stage Tandetron accelerator, which produces low energy neutrons via the 7Li(p, n)7Be reaction. The mean energy of the neutron beams produced ranged from 34 keV to 354 keV. The results presented in this study suggests that the compact METEPC is able to produce measurements in low dose rate radiation environments with the same precision in a given length of time as could be obtained with a 12.7 cm diameter spherical TEPC. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 13504487
- Volume :
- 45
- Issue :
- 10
- Database :
- Academic Search Index
- Journal :
- Radiation Measurements
- Publication Type :
- Academic Journal
- Accession number :
- 55626244
- Full Text :
- https://doi.org/10.1016/j.radmeas.2010.05.025