Quantitative monitoring of terbium ion by a Tb3+ selective electrode based on a new Schiff's base

Authors :
Zamani, Hassan Ali
Zabihi, Mohammad Saleh
Rohani, Mitra
Zangeneh-Asadabadi, Abbas
Ganjali, Mohammad Reza
Faridbod, Farnoush
Meghdadi, Soraia
Source :
Materials Science & Engineering: C. Mar2011, Vol. 31 Issue 2, p409-413. 5p.
Publication Year :
2011

Abstract

Abstract: Solution study showed N,N′-bis(5-nitrosalicylidene)-2-aminobenzylamin (L) trends toward Tb3+ ion. Then, it was used as a suitable ionophore in construction of terbium ion selective electrode. The electrode with composition of 30% PVC, 65% solvent mediator (NB), 3% ionophore (L) and 2% anionic additive (NaTPB) shows the best potentiometric response characteristics. It displays a Nernstian behavior (20.1mVdecade−1) over the concentration range 1.0×10−6 to 1.0×10−2 molL−1. The detection limit of the electrode is 6.3×10−7 molL−1. It has a very short response time (~10s) and a useful working pH range of 2.6–9.4 for at least 2months. The proposed membrane sensor shows excellent discriminating ability towards Tb3+ ions with regard to several alkali, alkaline earth, transition and heavy metal ions. To investigate the analytical applicability of the sensor, it was successfully applied to the determination of terbium in certified reference material. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
09284931
Volume :
31
Issue :
2
Database :
Academic Search Index
Journal :
Materials Science & Engineering: C
Publication Type :
Academic Journal
Accession number :
57076892