Back to Search Start Over

In vitro estimation of mean sound speed based on minimum average phase variance in medical ultrasound imaging

Authors :
Yoon, Changhan
Lee, Yuhwa
Chang, Jin Ho
Song, Tai-kyong
Yoo, Yangmo
Source :
Ultrasonics. Oct2011, Vol. 51 Issue 7, p795-802. 8p.
Publication Year :
2011

Abstract

Abstract: Effective receive beamforming in medical ultrasound imaging is important for enhancing spatial and contrast resolution. In current ultrasound receive beamforming, a constant sound speed (e.g., 1540m/s) is assumed. However, the variations of sound speed in soft tissues could introduce phase distortions, leading to degradation in spatial and contrast resolution. This degradation becomes even more severe in imaging fatty tissues (e.g., breast) and with obese patients. In this paper, a mean sound speed estimation method where phase variance of radio-frequency channel data in the region of interest is evaluated is presented for improving spatial and contrast resolution. The proposed estimation method was validated by the Field II simulation and the tissue mimicking phantom experiments. In the simulation, the sound speed of the medium was set to 1450m/s and the proposed method was capable of capturing this value correctly. From the phantom experiments, the −18-dB lateral resolution of the point target at 50mm obtained with the estimated mean sound speed was improved by a factor of 1.3, i.e., from 3.9mm to 2.9mm. The proposed estimation method also provides an improvement of 0.4 in the contrast-to-noise ratio, i.e., from 2.4 to 2.8. These results indicate that the proposed mean sound speed estimation method could enhance the spatial and contrast resolution in the medical ultrasound imaging systems. [Copyright &y& Elsevier]

Details

Language :
English
ISSN :
0041624X
Volume :
51
Issue :
7
Database :
Academic Search Index
Journal :
Ultrasonics
Publication Type :
Academic Journal
Accession number :
61172875
Full Text :
https://doi.org/10.1016/j.ultras.2011.03.007