Back to Search Start Over

Radiothérapie conformationnelle avec modulation d’intensité : analyse des résultats des contrôles précliniques, expérience du centre René-Gauducheau

Authors :
Chiavassa, S.
Brunet, G.
Gaudaire, S.
Munos-Llagostera, C.
Delpon, G.
Lisbona, A.
Source :
Cancer Radiothérapie. Jul2011, Vol. 15 Issue 4, p265-269. 5p.
Publication Year :
2011

Abstract

Abstract: Purpose: Systematic verifications of patient''s specific intensity-modulated radiation treatments are usually performed with absolute and relative measurements. The results constitute a database which allows the identification of potential systematic errors. Material and methods: We analyzed 1270 beams distributed in 232 treatment plans. Step-and-shoot intensity-modulated radiation treatments were performed with a Clinac (6 and 23 MV) and sliding window intensity-modulated radiation treatments with a Novalis (6 MV). Results: The distributions obtained do not show systematic error and all the control meet specified tolerances. Conclusion: These results allow us to reduce controls specific patients for treatments performed under identical conditions (location, optimization and segmentation parameters of treatment planning system, etc.). [Copyright &y& Elsevier]

Details

Language :
French
ISSN :
12783218
Volume :
15
Issue :
4
Database :
Academic Search Index
Journal :
Cancer Radiothérapie
Publication Type :
Academic Journal
Accession number :
62670802
Full Text :
https://doi.org/10.1016/j.canrad.2010.10.005