Back to Search Start Over

Zincon-immobilized silica-coated magnetic Fe3O4 nanoparticles for solid-phase extraction and determination of trace lead in natural and drinking waters by graphite furnace atomic absorption spectrometry

Authors :
Jiang, Hong-mei
Yan, Zhong-peng
Zhao, Yue
Hu, Xin
Lian, Hong-zhen
Source :
Talanta. 5/30/2012, Vol. 94, p251-256. 6p.
Publication Year :
2012

Abstract

Abstract: A new protocol using zincon-immobilized silica-coated magnetic Fe3O4 nanoparticles (Zincon–Si–MNPs) as solid-phase extraction (SPE) medium has been developed for the separation and preconcentration of trace lead in water. Various parameters such as pH, extraction time, concentration and volume of eluent, sample volume, and influence of co-existing ions have been investigated in order to establish the optimum conditions for the determination of lead in combination with graphite furnace atomic absorption spectrometry (GFAAS). The detection limit (LOD) of the proposed method for lead based on an enrichment factor of 200 was 10ngL−1. The relative standard deviations (RSDs, n =5) were 8.3%, 7.8% and 9.2%, respectively, at 5, 0.5 and 0.05ngmL−1 levels. This method has been successfully applied to the analysis of trace lead in natural and drinking water samples and the recoveries for the spiked samples were in the range of 84–104%. [Copyright &y& Elsevier]

Details

Language :
English
ISSN :
00399140
Volume :
94
Database :
Academic Search Index
Journal :
Talanta
Publication Type :
Academic Journal
Accession number :
75449637
Full Text :
https://doi.org/10.1016/j.talanta.2012.03.035