Back to Search
Start Over
Flow-injection spectrophotometric determination of periodate and iodate by their reaction with pyrogallol red in acidic media
- Source :
-
Spectrochimica Acta Part A: Molecular & Biomolecular Spectroscopy . Nov2002, Vol. 58 Issue 13, p2835. 5p. - Publication Year :
- 2002
-
Abstract
- A simple and efficient flow-injection method was used with good results to determine periodate and iodate in water. The method is based on the reaction of periodate or iodate with pyrogallol red in sulfuric acid media. The reaction was monitored spectrophotometrically by measuring the decrease in absorbance of pyrogallol red at 470 nm. The chemical and FIA variables were established using the univariate and simplex optimization methods. The calibration curve was linear over the concentration ranges of 0.8–73.0 and 2.0–100.0 μM for IO4− and IO3−, respectively. The detection limits were 0.7 and 1.0 μM for IO4− and IO3−, respectively. The sample throughput was 20±5 h−1. The influence of potential interfering ions in the determination of the anions was tested. The relative standard deviations for 4.20 and 5.73 μM periodate and iodate were 1.8 and 2.2%, respectively. [ABSTRACT FROM AUTHOR]
- Subjects :
- *FLOW injection analysis
*IODATES
*ANIONS
Subjects
Details
- Language :
- English
- ISSN :
- 13861425
- Volume :
- 58
- Issue :
- 13
- Database :
- Academic Search Index
- Journal :
- Spectrochimica Acta Part A: Molecular & Biomolecular Spectroscopy
- Publication Type :
- Academic Journal
- Accession number :
- 7790405
- Full Text :
- https://doi.org/10.1016/S1386-1425(02)00032-X