Back to Search Start Over

Design and Analysis of MEMS Based PVDF Ultrasonic Transducers for Vascular Imaging.

Authors :
Chandrana, Chaitanya
Talman, James
Tao Pan
Roy, Shuvo
Fleischman, Aaron
Source :
Sensors (14248220). 2010, Vol. 10 Issue 8, p8740-8750. 11p.
Publication Year :
2010

Abstract

Polyvinilidene fluoride (PVDF) single-element transducers for high-frequency (>30 MHz) ultrasound imaging applications have been developed using MEMS (Micro-electro-Mechanical Systems) compatible techniques. Performance of these transducers has been investigated by analyzing the sources and effects of on-chip parasitic capacitances on the insertion-loss of the transducers. Modeling and experimental studies showed that on-chip parasitic capacitances degraded the performance of the transducers and an improved method of fabrication was suggested and new devices were built. New devices developed with minimal parasitic effects were shown to improve the performance significantly. A 1-mm aperture PVDF device developed with minimal parasitic effects has resulted in a reduction of insertion loss of 21 dB compared with devices fabricated using a previous method. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
14248220
Volume :
10
Issue :
8
Database :
Academic Search Index
Journal :
Sensors (14248220)
Publication Type :
Academic Journal
Accession number :
84942037
Full Text :
https://doi.org/10.3390/s100908740