Back to Search
Start Over
Characteristics of a CsI(Tl) Compton-suppressed clover detector.
- Source :
-
Journal of Physics G: Nuclear & Particle Physics . Mar2013, Vol. 40 Issue 3, p1-10. 10p. - Publication Year :
- 2013
-
Abstract
- The performance of a four fold segmented clover detector coupled to a CsI-suppressed shield has been tested using several standard radioactive sources. The summing characteristics, hit patterns and absolute efficiency were measured in both crystal and clover modes. Peak-to-total ratios of 57.3% and 45.6% in suppressed clover mode have been determined for 137Cs and 60Co sources, respectively. A Geant4 simulation of the suppressed clover detector, including the segmentation of the germanium crystals, has been created, and the simulation results agreed very well with the experimental data. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 09543899
- Volume :
- 40
- Issue :
- 3
- Database :
- Academic Search Index
- Journal :
- Journal of Physics G: Nuclear & Particle Physics
- Publication Type :
- Academic Journal
- Accession number :
- 85951179
- Full Text :
- https://doi.org/10.1088/0954-3899/40/3/035101