Back to Search Start Over

Simultaneous Determination of Hydroquinone, Catechol and Resorcinol at Graphene Doped Carbon Ionic Liquid Electrode.

Authors :
Li Ma
Guang-Chao Zhao
Source :
International Journal of Electrochemistry. 2012, p1-8. 8p. 4 Charts, 5 Graphs.
Publication Year :
2012

Abstract

new composite electrode has been prepared with doping graphene into the paste consisting graphite and ionic liquid, n-octyl-pyridinum hexafluorophosphate (OPFP). This electrode shows an excellent electrochemical activity for the redox of hydroquinone (HQ), catechol (CC), and resorcinol (RS). In comparison with bare paste electrode, the redox peaks of three isomers of dihydroxybenzene can be obviously, simultaneously observed at graphene doping paste electrode. Under the optimized condition, the simultaneous determination of HQ, CC, and RS in their ternary mixture can be carried out with a differential pulse voltammetric technique. The peak currents are linear to the concentration of HQ, CC, and RS in the range form 1 × 10-5 to 4 × 10-4, 1 × 10-5 to 3 × 10-4, and 1 × 10-6 to 1.7 × 10-4 mol L-1, respectively. The limits of detection are 1.8 × 10-6 mol L-1 for HQ, 7.4 × 10-7 mol L-1 for CC, and 3.6 × 10-7 M for RS, respectively. [ABSTRACT FROM AUTHOR]