Motion-Compensated Mega-Voltage Cone Beam CT Using the Deformation Derived Directly From 2D Projection Images.

Authors :
Chen, Mingqing
Cao, Kunlin
Zheng, Yefeng
Siochi, R. Alfredo C.
Source :
IEEE Transactions on Medical Imaging. Aug2013, Vol. 32 Issue 8, p1365-1375. 11p.
Publication Year :
2013

Abstract

This paper presents a novel method for respiratory motion compensated reconstruction for cone beam computed tomography (CBCT). The reconstruction is based on a time sequence of motion vector fields, which is generated by a dynamic geometrical object shape model. The dynamic model is extracted from the 2D projection images of the CBCT. The process of the motion extraction is converted into an optimal 3D multiple interrelated surface detection problem, which can be solved by computing a maximum flow in a 4D directed graph. The method was tested on 12 mega-voltage (MV) CBCT scans from three patients. Two sets of motion-artifact-free 3D volumes, full exhale (FE) and full inhale (FI) phases, were reconstructed for each daily scan. The reconstruction was compared with three other motion-compensated approaches based on quantification accuracy of motion and size. Contrast-to-noise ratio (CNR) was also quantified for image quality. The proposed approach has the best overall performance, with a relative tumor volume quantification error of 3.39 \pm 3.64% and 8.57 \pm 8.31% for FE and FI phases, respectively. The CNR near the tumor area is 3.85 \pm 0.42 (FE) and 3.58 \pm 3.33 (FI). These results show the clinical feasibility to use the proposed method to reconstruct motion-artifact-free MVCBCT volumes. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
02780062
Volume :
32
Issue :
8
Database :
Academic Search Index
Journal :
IEEE Transactions on Medical Imaging
Publication Type :
Academic Journal
Accession number :
89549810
Full Text :
https://doi.org/10.1109/TMI.2012.2231694